Disertación de Maestría
DOI
https://doi.org/10.11606/D.85.2014.tde-14112014-142426
Documento
Autor
Nombre completo
Paulo Marcio de Oliveira
Dirección Electrónica
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
São Paulo, 2014
Director
Tribunal
Rodrigues, Letícia Lucente Campos (Presidente)
Rocha, Felícia Del Gallo
Vivolo, Vitor
Título en portugués
Avaliação de dose nos sistemas de irradiação de bolsas de sangue para esterilização
Palabras clave en portugués
dosimetria química
dosímetro Fricke Gel
irradiação de bolsas de sangue
irradiação de hemocomponentes do sangue
Resumen en portugués
O procedimento de esterilização do sangue é adotado para que sejam mínimas as condições necessárias de manifestação da doença do enxertoversus- hospedeiro associada à transfusão de sangue (DECHT), síndrome, geralmente fatal. Neste trabalho foi utilizada a solução Fricke gel desenvolvida no IPEN para a obtenção de um objeto simulador de bolsa de sangue empregada no processo de irradiação para esterilização de sangue em um irradiador Gammacell Elan. O objetivo desse trabalho é determinar a homogeneidade da dose aplicada no processo de esterilização de bolsas de sangue. Foram utilizadas as técnicas de Espectrofotometria de Absorção Óptica (EAO), Densitometria Óptica (DO), Termoluminescência (TL) e Ressonância Magnética por Imagem (RMI) e os softwares Numaris® versão 4 (Syngo-MR-A30) e MatLab® (versão 9.0) para a aquisição e reconstrução das imagens. O estudo foi realizado posicionando no canister do irradiador inicialmente uma bolsa e, posteriormente, duas bolsas e três bolsas e aplicando uma dose nominal de 25 Gy. O fabricante do irradiador Gammacell Elan recomenda que seja irradiado em cada procedimento o valor máximo de três bolsas de sangue. Os resultados obtidos indicam que o irradiador Gammacell Elan não demostrou homogeneidade na distribuição da dose. Irradiando uma bolsa de sangue foi obtida uma diferença de dose de 3,65 %, enquanto para duas bolsas de sangue essa diferença caiu para 2,65 % e quando posicionadas três bolsas de sangue a diferença de dose foi de 1,88 %. O tempo de abertura de fechamento do equipamento de irradiação é responsável pela diferença de dose 3,17 % acima da dose prevista à realização das irradiações de bolsas de sangue. Os resultados obtidos usando o filme radiocrômico Grafchromic MD-55 de dupla face aplicada à avaliação de dose no sistema de rotação de irradiação não apresentou resultados reproduzíveis.
Título en inglés
Evaluation of radiation dose in systems of blood bag for sterilization
Palabras clave en inglés
chemical dosimetry
Fricke Gel dosimeter
irradiation of blood bags
irradiation of blood components from the blood
Resumen en inglés
The blood sterilization procedure is adopted so that the necessary conditions of manifestation of graft-versus-host syndrome, disease associated with blood transfusion (AT-GVHD), usually a fatal disease, are minimal. In this work the Fricke gel solution developed in IPEN was used for obtaining a phantom of blood bag used in the irradiation process for sterilization of blood in a Gammacell irradiator Elan. The objective of this study was to evaluate the homogeneity of the applied radiation sterilization dose in the blood bags. Optical Absorption Spectroscopy (OAS), Optical Densitometry (OD), Thermoluminescence (TL) and Magnetic Resonance Imaging (MRI) technics and Numaris® software version 4 (Syngo-MR-A30) and MatLab® (version 9.0) softwares were used to images acquisition and reconstruction. The study was conducted positioning one bag in the canister irradiator initially and subsequently two and three bags together and applying a nominal dose of 25 Gy. The Elan Gammacell irradiator manufacturer recommends that the maximum values of three blood bags are irradiated in each procedure. The results indicate that the Gammacell irradiator Elan did not presents homogeneous dose distribution. Radiating a blood bag a dose difference of 3.65 % was obtained, while for two bags of blood that difference dropped to 2.65% and when positioned three blood bags the difference in dose was 1.88 %. The opening and closing times of the irradiation source is responsible for an increasing in the planned irradiation dose of 3.17%. The results obtained using radiochromic double-sided Grafchromic MD-55 film applied to dose evaluation in the rotation irradiation system did not present reproducible results.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Fecha de Publicación
2014-12-22
 