Mémoire de Maîtrise
DOI
https://doi.org/10.11606/D.85.2012.tde-18022014-154354
Document
Auteur
Nom complet
Vinicius Demanboro Gonçalves
Adresse Mail
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
São Paulo, 2012
Directeur
Jury
Yoriyaz, Hélio (Président)
Fernandes, Marco Antonio Rodrigues
Scaff, Luiz Alberto Malaguti
Titre en portugais
Caracterização da dose em pacientes devido à produção de imagens de raios-X utilizadas em radioterapia guiada por imagem - IGRT
Mots-clés en portugais
dose
radioterapia
simulação Monte Carlo
Resumé en portugais
O processo de radioterapia consiste em várias etapas, iniciando na indicação pelo médico. O plano de tratamento passa então por um processo denominado simulação, onde é adquirida uma série de imagens por tomografia computadorizada que são transferidas para o sistema de planejamento, onde a delineação dos volumes alvos e tecidos normais adjacentes serão realizadas. Após a delineação desses volumes, no sistema de planejamento são colocados os campos de irradiação e a dose desejada conforme prescrição médica. O sistema de planejamento calcula então a dose que o volume alvo e os tecidos adjacentes poderão receber. Se estas doses estão dentro dos padrões aceitáveis, o planejamento então é aprovado e enviado ao acelerador linear para a execução do tratamento. Antes da execução do tratamento, é realizada uma imagem, seja através de filme radiográfico ou digitalmente, para avaliar a posição no paciente na mesa de tratamento. Se a localização do paciente está correta, a dose é então liberada. Esse protocolo de aquisição de imagem é denominado como Radioterapia Guiada por Imagem (IGRT). A quantidade de radiografias de posicionamento segue um protocolo definido conforme a região a ser irradiada. Como resultado deste procedimento, sabe-se que uma determinada dose adicional é recebida pelos pacientes, tornando-se um fator importante a ser determinado. Esta avaliação foi realizada através da simulação de Monte Carlo, utilizando o código MCNP. Para isso foi realizada primeiramente toda a caracterização da fonte de raios X com uso de câmaras de ionização e dosimetros TL juntamente com as simulações no MCNP. Após essa caracterização, as imagens e as estruturas do planejamento radioterápico foram convertidas para serem utilizadas no código MCNP. Para que as doses fossem calculadas nos principais órgãos de risco no tratamento de próstata: bexiga, reto e cabeças de fêmur direita e esquerda.
Titre en anglais
Characterization of dose in patients due to production of X-ray images used in image-guided radiotherapy - IGRT
Mots-clés en anglais
dose
Monte Carlo simulation
radiotherapy
Resumé en anglais
The process of radiotherapy treatment consists of several stages, starting from the statement given by the physician. The treatment planning undergoes a process called simulation, where a series of computed tomography images is acquired and transferred to the treatment planning system, where the delineation of target volumes and adjacent normal tissues will be performed. After the delineation of these volumes, then irradiation fields and dose precribed by the physician are placed in the treatment planning system. It calculates the dose that target volume and surrounding tissues are receiving. If the doses are within acceptable standard values, then the design is approved and submitted to the linear accelerator for the treatment course. Before treatment course, an image is performed, either by radiographic or digital film, in order to evaluate (check) the patient position on the treatment table. If the patient position is correct, the treatment is realized. This image acquisition protocol is called Image-Guided Radiotherapy (IGRT). The amount of radiographic positioning follow a protocol defined for the region to be treated. As a result of this procedure, it is known that a specific additional dose is received by the patient, becoming an important factor to be determined. In this work, this additional dose evaluation was performed by the Monte Carlo simulation using the MCNP algorithm. The characterization of the entire X-ray source was primarily realized with ionization chamber thermoluminescent dosimeters and simulations with the MCNP code. After the X-ray tube characterization, images and the structures for the radiotherapy planning were converted to be used in the MCNP code for dose calculation at the main organs at risk during a prostate treatment: bladder, rectum and femoral heads right and left.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
Date de Publication
2014-04-07
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.