Master's Dissertation
DOI
https://doi.org/10.11606/D.11.1978.tde-20240301-154721
Document
Author
Full name
Boaventura Freire dos Reis
Institute/School/College
Date of Defense
Published
Piracicaba, 1978
Supervisor
Title in Portuguese
Determinação colorimétrica de alumínio em águas naturais, plantas e solos por injeção em fluxo contínuo
Keywords in Portuguese
ÁGUAS NATURAIS
ALUMÍNIO
ANÁLISE POR INJEÇÃO EM FLUXO CONTÍNUO
COLORIMETRIA
DETERMINAÇÃO
PLANTAS
SOLOS
Abstract in Portuguese
Desenvolveu-se um método para a determinação de alumínio em extratos de solos, material vegetal pré-digerido e em águas naturais empregando o reagente colorimétrico Eriocromo cianina R e a metodologia de injeção em fluxo contínuo. Estabeleceu-se o sistema de fluxo de modo a se conseguir uma taxa de determinações de 100 a 120 amostras por hora. O consumo de reagentes é otimizado, para a determinação de alumínio no intervalo de 0,25 a 10,00 ppm. Eliminou-se a interferência do íon férrico durante a residência da amostra no percurso analítico, mascarando-o com ácido ascórbico ou com ácido tioglicólico. Verificaram-se os níveis de interferências do fosfato e do fluoreto, e a partir de que concentrações as respectivas interferências são significativas. Os consumos de Eriocromo cianina R, ácido ascórbico o ácido tioglicólico e de acetato de amônia são respectivamente 6%, 11% e 13% em relação ao método manual. A média de recuperação pelo teste de adição controlada é de 96%, e a reprodutibilidade é da ordem de 1%.
Abstract in English
The colorimetric determination of aluminium in soil extracts, plant digests and natural waters with Eriocrom cyanin R was adapted to continuous flow-injection analysis. The proposed system presents a typical sampling rate of about 120 samples per hour, when the aluminium content is in the range between 0,25 and 10,00 ppm Good reproducibility and accuracy associated with low consumption of sample and reagent, which are characteristics of the flow-injection approach, were also attained. The iron (III) interference was studied and suppressed after sample injection, by using ascorbic acid or tioglicolic acid as masking agents. The phosphate and fluoride interferences in the Al-determination by the proposed method are discussed.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
724639.pdf (2.47 Mbytes)
Publishing Date
2024-03-14
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.