• JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
 
  Bookmark and Share
 
 
Mémoire de Maîtrise
DOI
https://doi.org/10.11606/D.14.2006.tde-31052006-085228
Document
Auteur
Nom complet
Dionisio Uendro Carlos
Adresse Mail
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
São Paulo, 2006
Directeur
Jury
Ribeiro, Fernando Brenha (Président)
Shukowsky, Wladimir
Yoshimura, Elisabeth Mateus
Titre en portugais
"Desenvolvimento e construção de blocos de concreto radioativo para a calibração de espectrômetros gama portáteis e aerotransportados, utilizados em exploração mineral e de hidrocarbonetos"
Mots-clés en portugais
aerogamaespectrometria
blocos transportáveis de calibração
calibração
correção de stripping
levantamento aéreo
raios gama
Resumé en portugais
Oito blocos transportáveis para calibração foram construídos para serem utilizados como padrões de concentração na calibração de espectrômetros gama portáteis e aerotransportados. Todo o procedimento de construção é descrito em detalhe. Os blocos, com dimensões de 1 m x 1 m x 0,30 m e massas variando entre 593 kg e 673 kg, devem a sua radioatividade à adição de diferentes quantidades de feldspato potássico (ortoclásio), caldasito e areia monazítica à massa de concreto. As concentrações de potássio, urânio e tório variam significativamente de um bloco para outro atingindo valores máximos de 5,7% de K, 45,6 ppm eU e 137 ppm eTh. A distribuição do fluxo de radiação gama proveniente da superfície dos blocos de concreto e a magnitude das heterogeneidades na concentração dos elementos radioativos foi determinada experimentalmente. Como exemplo de aplicação apresenta-se o resultado da calibração de um espectrômetro gama portátil.
Titre en anglais
Development and Building of Radioactive Concrete Pads for calibration of the airborne and ground gamma-ray spectrometers, used in mineral exploration and hydrocarbons
Mots-clés en anglais
airborne gamma-ray spectrometry
calibration
gamma-ray
stripping correction
transportable calibration pads
Resumé en anglais
Eight transportable calibration pads were built in to be used as concentration standards for portable and airborne gamma spectrometers calibrations. The pads construction procedure is described in full detail. The pads, with dimensions of 1 m x 1 m x 0,30 m and masses between 593 kg and 673 kg were made radioactive by the addition of different amounts of k-feldspar, caldasite and monazitic sand to the concrete masses. The potassium, uranium and thorium concentration vary significantly in the pads, reaching maximum values of 5,7% of K, 45,6 ppm eU and 137 ppm eTh. The distribution of the gamma radiation flux from the pads surfaces and the heterogeneity magnitudes of the radioactive elements concentration were experimentally established. An example of gamma spectrometer calibration is presented.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
Date de Publication
2006-09-14
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.