Mémoire de Maîtrise
DOI
https://doi.org/10.11606/D.23.2005.tde-11112005-125541
Document
Auteur
Nom complet
Lilian Hitomi Nakamura
Adresse Mail
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
São Paulo, 2005
Directeur
Jury
Contin, Ivo (Président)
Adachi, Lena Katekawa
Tortamano Neto, Pedro
Titre en portugais
"Estudo comparativo do afrouxamento de diferentes parafusos de fixação de abutment em implantes de hexágono externo e interno, após o ensaio de ciclagem mecânica"
Mots-clés en portugais
Implante
Abutment
Ciclagem mecânica
Parafuso de fixação
Prótese dentária
Torque
Resumé en portugais
A proposta deste estudo foi avaliar o torque de afrouxamento de três tipos de parafusos de fixação de abutment. Foram formados seis grupos: grupos HE-TiS e HE-TiNS (implante e abutment de hexágono externo, parafuso de fixação de titânio, submetidos e não-submetidos ao ensaio), grupos HE-AuS e HE-Au NS (implante e abutment de hexágono externo, parafuso de fixação de ouro, submetidos e não-submetidos ao ensaio) e grupos HI-TiS e HI-TiNS (implante e abutment de hexágono interno, parafuso de fixação de titânio, submetidos e não-submetidos ao ensaio). Para a realização da ciclagem mecânica, foi desenvolvida uma máquina de ensaio cuja carga máxima aplicada foi de 120 N em sentido axial, com deslizamento de 0,3mm sobre a superfície oclusal da coroa e força de atrito de 28 N a uma freqüência de 60 ciclos por minuto , durante 500.000 ciclos. O torque de apertamento inicial dos parafusos de fixação foi de 32 Ncm e o de afrouxamento mensurado a cada 100.000 ciclos nos grupos submetidos ao ensaio de ciclagem mecânica e a cada 24 horas após o torque de apertamento nos grupos não-submetidos ao ensaio. Dentro das condições experimentais deste estudo, os resultados obtidos sugerem que: os valores de destorque dos parafusos de fixação foram menores em todos os grupos submetidos ao ensaio; o destorque foi estatisticamente menor para os parafusos de titânio quando comparados os grupos HE-TiS (15,56±1,38 Ncm – 48%) / HE-AuS (18,15±1,85 Ncm – 56%) e os grupos HE-TiNS (23,76±1,59 Ncm – 74%) / HE-AuNS (27,06±1,23 Ncm – 84,5%); o destorque dos parafusos de titânio também foi estatisticamente menor quando utilizados nos grupos HE-TiS (15,56±1,38 Ncm – 48%), comparados aos grupos HI-TiS (18,56±1,14 Ncm – 58%).
Titre en anglais
Comparative study of screw loosening for different abutment screw in external and internal hexagonal implant system after cyclic loading
Mots-clés en anglais
Abutment screw
Ciclyc loading
Denture prosthese
Implant
Torque
Resumé en anglais
The purpose of this study was to evaluate the abutment screw loosening when either submitted or not to cyclic loading. Six groups were composed: HE-TiS and HE-TiNS (external hexagonal implant and abutment, titanium abutment screw, submitted and not subimitted to cyclic loading), HE-AuS and HE-Au NS (external hexagonal implant and abutment, gold abutment screw, submitted not submitted to ciclyc loading), HI-TiS and HI-TiNS (internal hexagonal implant and abutment, titanium abutment screw, submitted and not submitted to ciclyc loading). In order to perform the cyclic loading, it was developed a device wich applied the maximum load of 120 N in the axial direction, with a slip of 0,3 mm on the crown oclusal surface with friction force of 28 N. The frequency was 60 cycles per minute for a duty of 500,000 cycles. The abutment screw was tightened to 32 Ncm and the loosening torque was measured after every 100,000 cycles in the submitted groups and every 24 hours after the tightening torque in the not-submitted groups. Within the limitations of this study, the results suggest that the loosening torque values were statistically significant lower in all groups submitted to the cyclic loading; the loosening torque was statiscally lower for titanium screws when compared HE-TiS (15,56±1,38 Ncm – 48%) / HE-AuS (18,15±1,85 Ncm – 56%) groups and HE-TiNS (23,76±1,59 Ncm – 74%) / HE-AuNS (27,06±1,23 Ncm – 84,5%) groups; the loosenig torque was also lower for the titanium screws when used to in the groups HE-TiS (15,56±1,38 Ncm – 48%) compared to HI-TiS (18,56±1,14 Ncm – 58%).
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
LilianHNakamura.pdf (582.48 Kbytes)
Date de Publication
2006-01-16
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.