José Raul Girondi
Dirección Electrónica
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
São Paulo, 2006
Director
Tribunal
Panella, Jurandyr (Presidente)
Magalhães, José Cássio de Almeida
Pereira, Marlene Fenyo Soeiro de Matos
Pinto, Roberto Heitzmann Rodrigues
Vita, Renato
Título en portugués
"Validação das ferramentas de calibragem de medidas em três sistemas de radiografia digital"
Palabras clave en portugués
Radiografia digital
Sensor CCD
Sensor placa de fósforo Odontometria
Resumen en portugués
O desenvolvimento tecnológico busca fatores que possibilitem uma menor exposição do paciente e profissional aos raios ionizantes, maior agilidade no atendimento ao paciente e maior auxilio no desenvolvimento do diagnóstico final. Assim a associação entre informática e radiologia básica, traduzidas como radiologia digital, fez-se conseguir os fatores buscados. Uma ferramenta dos sistema digitais é a calibração prévia às medidas. Com essas observações avaliamos tais ferramentas de três sistemas de radiografias digitais, objetivando validar e comprovar sua eficácia nas metodologias dos trabalhos presentes e futuros. Para tanto utilizamos 20 dentes humanos secos provindos do Banco de Dentes Permanentes da FOUSP, os quais foram radiografados com o auxílio de três sistemas de radiografia digital, direto e semidireto (Digora, RVG Trophy e Visualix). O padrão ouro das medidas dos dentes foi a verificação das dimensões dos dentes por meio de um paquímetro digital e posteriormente comparados os seus padrões de medidas com os softwares dos respectivos sistema. Os resultados foram analisados estatisticamente pelo ANOVA (métodos de Tukey e Dunnett), chegando às seguintes conclusões: existe diferença significante entre os aparelhos p<0,001; a média das medidas obtidas houve diferença significante (p<0,001) com o aparelho Digora foi menor que as médias obtidas com os aparelhos das marcas Visualix e RVG; considerando-se o padrão A e o padrão B de calibragem, não existe diferença estatística significante entre os avaliadores; considerando-se a calibragem sem padrão, um dos avaliadores obteve média mais elevada que os demais avaliadores no caso das medidas realizadas com o aparelho Digora, porém este fato não se repetiu quando da utilização dos outros aparelhos; comparando-se as medidas de referência (paquímetro) observa-se que mais uma vez o aparelho Digora (padrão A, padrão B e sem padrão) apresentou diferença significativa; os aparelhos das marcas Visualix e RVG não mostraram diferenças significativas quando comparados com o valor de referência.
Título en inglés
Assessment of calibration tools of three digital radiography systems
Palabras clave en inglés
CCD sensor
Digital radiography
Odontometry
phosphor plate system
Resumen en inglés
Technological development aims at reducing exposure of both the professional and the patient to ionizing radiation, improving patient treatment and aiding final diagnosis. The combination of computer technology and radiology, i.e. digital radiography, has made this possible. One of the tools of digital radiography systems is calibration prior to measurement. The purpose of this study was to assess the calibration tools of three digital radiography systems to verify their efficacy. Twenty human teeth obtained from the Human Teeth Bank of the School of Dentistry, University of São Paulo, were used. The dry teeth were radiographed using three different direct digital radiography systems Digora, RVG Trophy and Visualix. The measurements of the teeth obtained with a digital caliper were considered the gold standard. These measurements were later compared with those obtained using the digital systems. The results were submitted to ANOVA (methods of Tukey and Dunnett), and the following conclusions were drawn: there was a significant difference between the systems used (p < 0.001); the averages of the measurements obtained with the three systems were significantly different (p<0,001). Visualix and RVG were more accurate than Digora considering A and B calibration standards, no significant difference was observed between the examiners; when there was no calibration standard and the Digora system was used, one examiner obtained more high averages measurements than the others; when there was no calibration standard and Visualix or RVG systems were used, no differences were observed between the examiners; the measurements obtained with Digora system (standards A, B or no standard) were statistically different from the gold standard the measurements obtained with Visualix and RVG systems presented no significant difference when compared with the gold standard.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
JoseRaulGirondi.pdf (591.96 Kbytes)
Fecha de Publicación
2006-05-18
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.