Thèse de Doctorat
DOI
https://doi.org/10.11606/T.3.2018.tde-18072018-135559
Document
Auteur
Nom complet
Raphael Garcia Moreira
Adresse Mail
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
São Paulo, 2018
Directeur
Jury
Santos Filho, Sebastião Gomes dos (Président)
Mansano, Ronaldo Domingues
Santos, Edson Tafeli Carneiro dos
Silva, Maria Lucia Pereira da
Siqueira, Cesareo de La Rosa
Titre en portugais
Fabricação e modelagem de uma nova geometria para espectrometria de mobilidade iônica de tolueno, propanol e água com elevada relação sinal-ruído.
Mots-clés en portugais
Descarga elétrica
Espectrometria
Espectrômetro
Ionização de gases
Relação entre sinal e ruído
Resumé en portugais
Neste trabalho é apresentada uma nova geometria e os critérios de projeto para dispositivos de espectrometria de mobilidade iônica, com objetivo de aumentar a relação entre sinal e ruído detectados pelos eletrômetros que compõem a região de detecção, quando associado à ionização baseada em descargas elétricas que tipicamente emitem maior interferência eletromagnética em relação a outros tipos de ionização. A geometria e os critérios de projeto foram estabelecidos com o auxílio do simulador de trajetórias iônicas SIMION versão 8.1 e comparada com espectrômetros de mobilidade iônica do estado-da-arte. Os resultados das simulações dos dispositivos foram avaliados por meio dos testes estatísticos não-paramétricos de Iman-Davenport e Holm. O resultado indicou que a geometria proposta apresentou compatibilidade com as melhores características dos espectrômetros atuais com confiança estatística a 95 %. Na sequência a geometria proposta foi simulada, fabricada e testada para detecção dos íons de: tolueno - C7H8 (C5H5+, C6H5+, C7H7+), propanol - C3H8O (CH3+, CH3O+, C2H3+, C2H4O+, C2H5O+, C3H6+, C3H7+, C3H7O+) e água - H2O (H+, HO+, H2+, O+). Os picos de corrente iônica detectada nos eletrômetros indicaram coerência com as trajetórias simuladas para estes conjuntos de íons. Além disso, os resultados experimentais apresentaram um aumento na relação entre sinal e ruído para a nova geometria proposta que foi validada pelo teste de comparação pareada não-paramétrico de Wilcoxon a 95 % de confiança estatística.
Titre en anglais
Fabrication and modeling of a new geometry for ion mobility spectrometry of toluene propanol and water with high signal-to-noise ratio.
Mots-clés en anglais
Discharge ionization
Ion mobility spectrometry
Signal-to-noise ratio
Spectrometer
Resumé en anglais
In this work, a new geometry and a design criteria for ion mobility spectrometry devices was presented, aiming to increase the signal-to-noise ratio detected by electrodes of the detection region, when associated with ionization based on electrical discharges. The new geometry and the design criteria were established in the SIMION version 8.1 (an ionic trajectory simulator) and compared with the state-of-art ion mobility spectrometers. The results from device simulations were evaluated using the non-parametric statistical test of Iman-Davenport and Holm. The result indicated that the proposed design criteria presented compatibility with the best characteristics of current spectrometers with statistical evidence at 95%. The proposed geometry was simulated, fabricated and tested for the following solvents: toluene - C7H8 (C5H5+, C6H5+, C7H7+), propanol - C3H8O (CH3+, CH3O+, C2H3+, C2H4O+, C2H5O+, C3H6+, C3H7+, C3H7O+) and water - H2O (H+, HO+, H2+, O+). The peaks of ion current detected in the electrometers occurred at approximate positions of the simulated trajectories for these sets of ions. In addition, the experimental data showed an increase in the signal-to-noise ratio for the proposed new geometry that was validated by the test statistic of Wilcoxon signed rank at 95% statistical confidence.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
Date de Publication
2018-07-26
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.