Problemas de segurança e desempenho em equipamentos eletromédicos associados aos diferentes padrões de alimentação elétrica.
Mots-clés en portugais
engenharia biomédica
engenharia elétrica
segurança de equipamentos
Resumé en portugais
Alguns problemas encontrados na certificação de equipamentos eletromédicos relacionados aos diferentes padrões de alimentação elétrica existentes no Brasil e em outros países são apresentados. Para analisá-los foi executado um conjunto de ensaios em um grupo de equipamentos eletromédicos operando em diferentes padrões de alimentação elétrica. Resultados com diferenças bastante significativas entre as cinco condições de alimentação utilizadas foram obtidos, concluindo-se que há necessidade de uma adaptação normativa que inclua prescrições e requisitos claros relativos às condições de alimentação elétrica nas quais os equipamentos devem ser ensaiados e certificados no Brasil, corrigindo algumas falhas normativas das publicações nacionais.
Titre en anglais
Safety and performance problems in medical electrical equipment related to different standards of mains supply.
Mots-clés en anglais
biomedical engineering
electrical engineering
equipment safety
Resumé en anglais
In this research were presented some problems encountered in the medical electrical equipment certification related to different mains supply standards existent in Brazil and in other countries. To analyse it, a set of tests in a group of medical electrical equipment operating in different mains supply was carried out. Results presented differences among the five used conditions of mains supply, indicating the necessity for a normative adaptation, including clear requirements related to the conditions of mains supply in which the equipment must be tested and certified in Brazil and correcting some national standard imperfections.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
alessandro.cavina.pdf (1.51 Mbytes)
Date de Publication
2005-05-17
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.