Mémoire de Maîtrise
DOI
https://doi.org/10.11606/D.3.2010.tde-19012011-094249
Document
Auteur
Nom complet
Alberto Mitsuo Oyama
Adresse Mail
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
São Paulo, 2010
Directeur
Jury
Itiki, Cinthia (Président)
Kohn, André Fábio
Manzano, Gilberto Mastrocola
Titre en portugais
Eliminação de artefatos de estímulo em potenciais evocados somatossensitivos.
Mots-clés en portugais
Ajuste de curvas
Artefato de estímulo
Potenciais evocados somatossensitivos
Resumé en portugais
Os potenciais evocados têm uma consagrada utilização em clínica. Sua obtenção é dificultada pela presença de outros sinais biológicos, de artefatos de movimento, de ruído eletrônico, de interferência da rede elétrica e de artefatos de estímulo. A média síncrona ou promediação é um método que elimina os sinais que não estejam sincronizados com a estimulação, incluindo os outros sinais biológicos, os artefatos de movimento, o ruído e a interferência. No entanto, esse método não consegue eliminar os artefatos de estímulo. Outros métodos devem ser usados para essa tarefa. Para esses métodos, a eliminação do artefato de estímulo é bem sucedida quando o artefato não se sobrepõe ao potencial evocado. Porém, para uma captação próxima ao local de estimulação, a sobreposição ocorre e dificulta a eliminação do artefato de estímulo. O objetivo deste trabalho foi o de estudar a variação da amplitude e latência do pico do potencial evocado e sua influência nas estimativas da amplitude, da latência e do erro quadrático médio. Para potenciais evocados em que houve sobreposição com o artefato, o erro médio quadrático sempre foi reduzido com a remoção do artefato de estímulo. O erro de medição da latência foi reduzido a praticamente zero, independentemente da amplitude do potencial evocado. Por outro lado, o método inseriu erro na medição da amplitude de potenciais evocados grandes. Por isso, nesse caso específico de atraso pequeno e amplitude grande, a medição da amplitude deve ser feita diretamente no sinal antes da remoção do artefato de estímulo. Comparando a ocorrência de sobreposição com os locais de captação do potencial evocado, pode-se afirmar que, para o modelo de artefato de estímulo usado neste trabalho, a necessidade de se aplicar o procedimento de remoção de artefato se restringiu aos potenciais evocados captados no cotovelo, para estimulação do nervo mediano tanto no punho quanto na mão.
Titre en anglais
Removal of stimulus artifact in somatosensory evoked potentials.
Mots-clés en anglais
Curve fitting
Somatosensory evoked potentials
Stimulus artifact
Resumé en anglais
Evoked potentials have been used in clinics. Their measurement is hindered by the presence of other biological signals, movement artifacts, electronic noise, power-line interference, and stimulus artifacts. Synchronous averaging is a method that eliminates the signals that are not synchronized with the stimulation, including other biological signals, movement artifacts, noise and interference. However, this method fails to eliminate stimulus artifacts. Other methods must be used in this task. Using these methods, one can obtain success in the stimulus artifact elimination, whenever the artifact does not superimpose with the evoked potential. Nevertheless, for a measurement close to the stimulation site, the superimposition is a fact that hinders the elimination of the stimulus artifact. The objective of this Masters thesis was to study the variation of the amplitude and latency of an evoked potential and verify their influence on the amplitude and latency estimates, as well as on the mean square error. For evoked potentials in which there was superposition, the mean square error was always reduced by the removal of the stimulus artifact. Latency measurement errors were reduced to zero, regardless of the evoked potential amplitude. However, this method inserted amplitude measurement errors for large evoked potentials. So, in the case of short delays and large amplitudes, amplitude measurements should be performed directly on the signal, before stimulus artifact removal. By comparing the presence of superposition with the evoked potential recording sites, one may state that, for the stimulus artifact model used in this work, the need to apply the artifact removal procedure was restricted to the evoked potentials recorded on the elbow, for median nerve stimulation both on wrist and hand.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
Date de Publication
2011-04-18
 
AVERTISSEMENT: Le matériau se réfère à des documents provenant de cette thèse ou mémoire. Le contenu de ces documents est la responsabilité de l'auteur de la thèse ou mémoire.
  • OYAMA, A M, and Itiki, C. Influence of evoked response latency and amplitude on stimulus artifact removal [doi:10.1109/IEMBS.2010.5627635]. In 2010 32nd Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC 2010), Buenos Aires. 2010 Annual International Conference of the IEEE Engineering in Medicine and Biology. : IEEE, 2010.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.