Doctoral Thesis
DOI
https://doi.org/10.11606/T.46.2004.tde-19052007-233213
Document
Author
Full name
Paulo Rogério Miranda Correia
E-mail
Institute/School/College
Knowledge Area
Date of Defense
Published
São Paulo, 2004
Supervisor
Committee
Oliveira, Pedro Vitoriano de (President)
Bertotti, Mauro
Krug, Francisco José
Pasquini, Célio
Rocha, Fábio Rodrigo Piovezani
Title in Portuguese
Estudo da utilização de padrão interno em determinações multielementares por espectrometria de absorção atômica com atomização eletrotérmica e detecção simultânea
Keywords in Portuguese
Análise de elementos-traço
Atomização eletrotérmica
Cádmio
Chumbo
Detecção simultânea
Determinação multielementar
Espectrometria de absorção atômica
Forno de grafite
Fuidos biológicos
Manganês
Níquel
Padronização interna
Sangue
Selênio
Soro sangüíneo
Urina
Abstract in Portuguese
Um estudo sistemático a respeito da utilização de padrão interno em determinações multielementares por espectrometria de absorção atômica (ETAAS) foi desenvolvido. O objetivo principal do presente trabalho foi verificar a possibilidade de melhorar a precisão e a exatidão dos resultados analíticos, que são obtidos na análise de fluidos biológicos. O pré-tratamento dessas amostras foi simplificado e reduzido a uma única etapa de diluição com surfactante (Triton X-100) e ácido (HNO3). Conseqüentemente, a complexidade da solução diluída de amostra, a ser introduzida no tubo de grafite, apresenta uma elevada quantidade de concomitantes que podem provocar interferências químicas. A seleção preliminar dos elementos a serem testados como padrão interno considerou a semelhança de parâmetros físico-químicos relacionados com o processo de atomização. Desta forma, Ag, Bi, In e Tl foram testados como padrão interno para a determinação simultânea de Cd/Pb em sangue e urina, enquanto Bi, Ge, In, Sb, Sn e Te foram os elementos selecionados para a determinação de Mn/Ni/Se em soro sangüíneo. A melhoria da qualidade dos resultados analíticos obtidos na determinação simultânea de Cd e Pb em sangue foi observada quando Ag foi utilizada como padrão interno, na presença de NH4H2PO4 como modificador químico. Verificou-se uma melhoria na exatidão dos resultados obtidos para Cd e Pb, após a correção com padrão interno. Por outro lado, os resultados obtidos na análise de urina não foram corrigidos por nenhum dos elementos testados. Os melhores resultados para a determinação simultânea de Mn, Ni e Se foram obtidos com a utilização de Bi, Sn e Te como padrão interno. Entretanto, verificou-se que a correção de todos os resultados não seria viável com o uso de um único padrão interno. O melhor desempenho nos testes realizados na presença de soro sangüíneo foi obtido com Bi, que melhorou discretamente a precisão dos resultados obtidos para Se. Desta forma, a padronização interna visando a determinação simultânea de Mn, Ni e Se não foi eficiente. A padronização interna em ETAAS, com a finalidade de melhorar a precisão e a exatidão dos resultados analíticos, é uma estratégia tão complexa, quanto os efeitos interferentes que se pretende corrigir: são necessários mais estudos para compreender melhor como a utilização de uma condição de compromisso afeta os processos de atomização, bem como mais informações a respeito das interferências físicas e químicas causadas por amostras complexas, analisadas por ETAAS após uma simples etapa de diluição. Deve-se considerar com especial atenção o modificador químico e as temperaturas das etapas de pirólise e de atomização empregadas, que são parâmetros críticos para o desempenho de um elemento como padrão interno.
Title in English
Study of the use of internal standard for multielement determinations by electrothermal atomic absorption spectrometry with simultaneous detection
Keywords in English
Atomic absorption spectrometry
Biological fluids
Blood
Cadmium
Electrothermal atomization
Graphite furnace
Internal standardization
Lead
Manganese
Multielement determination
Nickel
Selenium
Serum
Simultaneous detection
Trace-element analysis
Urine
Abstract in English
A systematic study involving the use of internal standard for multielement determinations by electrothermal atomic absorption spectrometry was developed. The main objective of this work was evaluate the possibility of improving precision and accuracy of the analytical results for biological fluids. The sample pre-treatment was reduced to a single dilution step with surfactant (Triton X-100) and acid (HNO3), increasing the amount of concomitant introduced into the atomizer. The preliminary selection of the elements to be tested as internal standard considered the resemblance of physico-chemical parameters related with the atomization process. Thus, Ag, Bi, In and Tl were tested as internal standard for the simultaneous determination of Cd/Pb in blood and urine, and Bi, Ge, In, Sb, Sn and Te were the selected elements for the determination of Mn/Ni/Se in blood serum. The correction of the results obtained for the simultaneous determination of Cd and Pb in blood was achieved when Ag was used as internal standard, in presence of NH4H2PO4 as chemical modifier. An improvement for the accuracy of the results was observed for both analytes after their correction with the internal standard. On the other hand, the results obtained for the urine analysis were not corrected by using the tested elements. The best results for the simultaneous determination of Mn, Ni and Se were observed when Bi, Sn and Te were used as internal standard. However, the correction for the results for all analytes was not possible by using only one internal standard. The best performance in presence of the serum was obtained for Bi, which improves slightly the precision for the Se results. Thus, the internal standardization for the simultaneous determination of Mn, Ni and Se was not efficient. The internal standardization in ETAAS, aiming the improvement of precision and accuracy of the analytical results, is a strategy as complex as the interference effects to be corrected: more studies are required in order to better understand how the adoption of a compromised condition disturbs the atomization processes, as well as to get more information about the physical and chemical interference caused by complex samples, analyzed by ETAAS after a single dilution step. The chemical modifier and the selected temperatures for the pyrolysis and atomization steps are critical parameters for the performace of an internal standard and they should be carefully considered.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
Tese.pdf (2.81 Mbytes)
Publishing Date
2007-10-19
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.