Author
Full name
Dosil Pereira de Jesus
E-mail
Institute/School/College
Knowledge Area
Date of Defense
Published
São Paulo, 1999
Supervisor
Committee
Lago, Claudimir Lúcio do (President)
Juliano, Valmir Fascio
Masini, Jorge César
Title in Portuguese
Cristais piezelétricos de quartzo com eletrodos separados e superfície modificada como sensores em fase líquida
Keywords in Portuguese
Análise por injeção em fluxo
Cristais piezelétricos de quartzo
Modificação química de superfície
Química analítica
Sensores
Abstract in Portuguese
Esta dissertação descreve a construção e avaliação de um detector para fase líquida utilizando cristal piezelétrico de quartzo (PQC) (3,579545MHz) com a configuração de eletrodos separados. A cela de detecção possui dois eletrodos de latão, que foram posicionados a uma pequena distância (cerca de alguns décimos de milímetros) da superfície do cristal, cujos eletrodos de prata originais foram removidos. Um dos espaços existentes entre os eletrodos e o PQC é preenchido com ar e o outro com um líquido, o qual pode fluir pela cela. Os eletrodos são conectados a um oscilador, cuja freqüência de oscilação é medida através de um freqüencímetro interfaceado a um microcomputador. O detector foi caracterizado, no que se refere à influência da condutividade, densidade, viscosidade e vazão da solução, que flui pela cela, na freqüência de oscilação do PQC. Foi estudado também, com o auxílio de um analisador de impedâncias, o comportamento do módulo e do ângulo de fase da impedância elétrica oferecida pela cela de detecção para a variação das propriedades reológicas, mencionadas anteriormente, da solução. Para testar o arranjo como sensor gravimétrico em um sistema de análise por injeção em fluxo (FIA), foram desenvolvidos dois métodos de silanização da superfície do cristal, um envolvendo o reagente trimetilclorosilano (TCS) e outro utilizando o composto N-(β-aminoetil)-γ-aminopropiltrimetoxisilano (AEAPTS). Os PQCs silanizados com o TCS foram testados para a detecção de compostos orgânicos (clorofórmio e acetona), enquanto os outros cristais, silanizados com AEAPTS, foram utilizados para detecção de íons metálicos. Para esses últimos PQCs notou-se uma boa sensibilidade para íons Cu2+, com os quais conseguiu-se um límite de detecção de aproximadamente 20µM.
Title in English
Piezoelectric quartz crystals with electrodes separated and modified surface as sensors in liquid media
Keywords in English
Analytical chemistry
Chemical surface modification
Flow injection analysis
Piezoelectric quartz crystal
Sensors
Abstract in English
This work describes the construction and evaluation of a liquid-phase detector using a piezoelectric quartz crystal (PQC) (3,579545 MHz) with separated electrodes. The detection cell has two brass electrodes, which were placed at a small distance (a tenth of millimeters) from the crystal surface, whose original silver electrodes were removed. One of the gaps between the crystal and the electrodes is filled with air, while the other one is filled with a liquid, which flows through the cell. The electrodes are connected to an oscillator, whose oscillation frequency is measured by a frequency counter interfaced to a microcomputer. The influence of the conductivity, density, viscosity and flow rate of the solution on the PQC oscillation frequency was determined. The behavior of the impedance of the cell and its phase angle as a function of the previously mentioned reologic properties ofthe solution were also studied using an impedance analyzer. Two methods for the silanization of the crystal surface, using trimethylchlorosilane (TCS) and N-(β-aminoethyl)-γ-aminopropyltrimethoxysilane (AEAPTS), were developed to test the assembly as a gravimetric sensor in a flow injection analysis system. The TCS-silanized PQCs were tested for the detection of organic compounds (chloroform and acetone), while the AEAPTS-silanized PQC were used for the detection of metallíc íons. These last PQC showed a good sensibility to Cu2+ ions, for which a limit of detection of about 20µM was obtained.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
Publishing Date
2009-06-25
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.