Mémoire de Maîtrise
DOI
https://doi.org/10.11606/D.46.2010.tde-25052011-142352
Document
Auteur
Nom complet
Kelliton José Mendonça Francisco
Adresse Mail
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
São Paulo, 2010
Directeur
Jury
Lago, Claudimir Lucio do (Président)
Gutz, Ivano Gebhardt Rolf
Oliveira, Marcone Augusto Leal de
Titre en portugais
Implementação e otimização de detector condutométrico sem contato para eletroforese capilar
Mots-clés en portugais
C4D
Cafeína
Condutometria
Detecção condutométrica sem contato capacitivamente acoplada
Eletroforese capilar
Faixa dinâmica
Ions alcalinos
Resumé en portugais
A presente dissertação trata da implementação e otimização de um sistema de detecção condutométrica sem contato capacitivamente acoplada (C4D) para Eletroforese Capilar (CE). O sistema é caracterizado pela compactação do sistema de detecção, versatilidade e flexibilidade de instalação em diferentes equipamentos comerciais de eletroforese capilar e home made. Desde a década de 80, a eletroforese capilar vem se consolidando como uma das técnicas de separação mais relevantes. Normalmente, os instrumentos comerciais são disponibilizados com detectores ópticos e detectores eletroquímicos. A C4D é utilizada em eletroforese capilar posicionando-se dois eletrodos tubulares envoltos ao capilar. A aplicação de sinais de alta frequência entre os eletrodos permite monitorar variações de condutividade da solução dentro do capilar. Assim, a resposta do detector depende de diversos fatores como mobilidade do analito, do co-íon do eletrólito, da frequência e amplitude do sinal aplicado entre os eletrodos e da geometria dos mesmos. A ausência de componentes móveis torna o presente detector compacto (6,5 cm3) e robusto. O presente C4D é constituído de um oscilador local funcionando a 1,1 MHz, um circuito capaz de converter corrente em tensão, retificar e filtrar os sinais analógcos, além de apresentar um conversor de sinais analógicos em digitais (Conversor A/D) de alta resolução (21 bits efetivos). Outra vantagem é a possibilidade de determinar o diâmetro interno do capilar através da leitura do conversor A/D quando se faz passar um fluxo de ar e, posteriormente água através do capilar. A diferença de leitura de conversor A/D pode ser convertida em diâmetro interno mediante uma curva de calibração. Em uma aplicação típica, os limites de detecção baseado no critério 3s (sem filtragem inicial) foi de 0,6, 0,4, 0,3, 0,5, 0,6 e 0,8 µmol L-1 para K+, Ba2+, Ca2+, Na+, Mg2+ e Li+, respectivamente, é comparável a outras implementações de alta qualidade de um detector condutométrico sem contato capacitivamente acoplado.
Titre en anglais
Implementation and optimization of contactless conductometric detector for capillary electrophoresis
Mots-clés en anglais
Alkali ions
C4D
Caffeine
Capacitively coupled contactless conductivity detector
Capillary electrophoresis
Conductometric
Dynamic range
Resumé en anglais
This dissertation deals with the implementation and optimization of a Capacitively Coupled Contactless Conductivity Detector (C4D) applied to capillary electrophoresis (CE). The system is characterized by compaction of the detection system, installation flexibility and versatility in different commercial capillary electrophoresis equipment and home-made. Since the 80s, capillary electrophoresis has been consolidated as one of the most important separation techniques. Typically, commercial instruments are available to optical detectors and electrochemical detectors. The C4D is used in capillary electrophoresis positioning two electrodes encased in tubular capillary. The application of high frequency signals between the electrodes allows monitoring changes in conductivity of the solution within the capillary. Thus, the detector response depends on several factors such as mobility of the analyte, the co-ion electrolyte, the frequency and amplitude of the signal applied between the electrodes and the geometry of the same. The absence of moving parts makes this compact detector (6.5 cm3) and robust. This C4D is composed of a local oscillator, working at 1.1 MHz, a circuit capable of converting current into voltage, rectify and filter the analogs signals, besides presenting a high-resolution analog-to-digital converter (ADC) (21 bits effective). Another advantage is the possibility of determining the internal diameter of the capillary by reading the ADC when it is passed a stream of air and then water through the capillary. A readings different ADC can be converted into inner diameter through a curve a calibration curve. In a typical application, the detection limits based on the 3s criterion (without baseline filtering) was 0.6, 0.4, 0.3, 0.5, 0.6 and 0.8 µmol L-1 for K+, Ba2 +, Ca2 +, Na+, Mg2+ and Li+, respectively, is comparable to other high-quality implementations of a capacitively coupled contactless conductivity detector.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
Date de Publication
2011-06-28
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.