Thèse de Doctorat
DOI
https://doi.org/10.11606/T.5.2005.tde-21122005-140229
Document
Auteur
Nom complet
Hussein Amin Orra
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
São Paulo, 2005
Directeur
Jury
Leao, Pedro Puech (Président)
Netto, Baptista Muraco
Silva, Domingos Guerino
Silva, Erasmo Simao da
Sitrângulo Junior, Cid José
Titre en portugais
"Estudo experimental da transmissão da pulsatilidade da endoprótese à parede do aneurisma da aorta após correção endoluminal"
Mots-clés en portugais
ANEURISMA AÓRTICO/cirurgia
CADÁVER
FLUXO PULSÁTIL
HUMANO
IMPLANTE DE PRÓTESE VASCULAR
PRÓTESE VASCULAR
Resumé en portugais
Objetivo: Medir a pulsatilidade da parede do aneurisma de aorta humano antes e depois de sua correção endoluminal. Método: Cinco aneurismas foram submetidos à perfusão pulsátil antes e depois do implante de uma endoprótese. Resultado: o nível da coluna de água oscilou durante a pulsação com variações de 17, 16, 13, 7 e 25 cm antes da colocação da endoprótese. Depois da prótese, a oscilação diminuiu em todos os casos para 13, 12, 9, 3,5 e 23 cm, respectivamente. Conclusão: A pulsação da endoprótese é transmitida à parede do aneurisma
Titre en anglais
Pulsatility transmission from endograft to aortic aneurysm wall after endovascular repair : an experimental study
Mots-clés en anglais
AORTIC ANEURYSM/surgery
BLOOD VESSEL PROSTHESIS
CADAVER
CAPILLARY PERMEABILITY
HUMAN
PROSTHESIS IMPLANTATION
PULSATILE FLOW
Resumé en anglais
Objective: To measure the pulsatility of human aortic aneurysms before and after exclusion with endograft. Method: Five aneurysms were submitted to pulsatile perfusion before and after implantation of a bifurcated endograft. Result: The level of the water column oscillated during pulsation, in each case, with an amplitude of 17, 16, 13, 7 and 25 cm before the endograft insertion. After that, the amplitudes dropped to, respectively 13, 12, 9, 3.5 and 23 cm.Conclusion: Pulsation of an endograft is transmitted to the aneurysm wall even in the absence of endoleak
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
HusseinAminOrra.pdf (696.88 Kbytes)
Date de Publication
2006-01-04
 