• JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
 
  Bookmark and Share
 
 
Tesis Doctoral
DOI
https://doi.org/10.11606/T.5.2019.tde-27022019-151124
Documento
Autor
Nombre completo
Bruno Francisco de Freitas Tonelotto
Dirección Electrónica
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
São Paulo, 2018
Director
Tribunal
Auler Junior, Jose Otavio Costa (Presidente)
Malbouisson, Luiz Marcelo Sá
Mathias, Ligia Andrade da Silva Telles
Neves, Flavio Humberto de Sousa
Título en portugués
Determinação da PEEP ideal e avaliação de atelectasia pulmonar com o uso da ultrassonografia durante intraoperatório de cirurgias eletivas
Palabras clave en portugués
Assistência perioperatória
Atelectasia pulmonar
Hiperinflação pulmonar
Linhas H
Manobra de recrutamento
Pulmão/diagnóstico por imagem
Respiração com pressão positiva
Ventilação pulmonar
Resumen en portugués
Introdução: A atelectasia intraoperatória ocorre imediatamente após a indução anestésica e pode ser detectada por ultrassom pulmonar (LUS). No entanto, até o momento o LUS não é utilizado para avaliar a hiperdistensão pulmonar. Neste estudo, descreveu-se um método para detectar hiperdistensão pulmonar usando LUS. A tomografia de impedância elétrica (TIE) foi a referência para comparação dos métodos. Métodos: Dezoito (18) pacientes, com 63 ± 6 anos de idade, com pulmões normais, submetidos à cirurgia abdominal inferior. O TIE foi calibrado, realizada a indução anestésica, intubação e ventilação mecânica. Para reverter a atelectasia posterior, realizou-se uma manobra de recrutamento alveolar com o uso de pressão expiratória final positiva (PEEP) 20 cmH20 e pressão aérea do platô 40 cmH2O durante 120 segundos. A titulação PEEP foi então obtida com valores descendentes: 20, 18, 16, 14,12,10, 8, 6 e 4 cmH2O. Os dados de ultrassom e TIE foram coletados em cada nível PEEP e interpretados por dois observadores independentes. O número de linhas H foi contado usando um filtro especial. O teste de correlação de Spearman e a curva ROC foram utilizados para comparar os dados do LUS e TIE. Resultados: O número de linhas H aumentou linearmente com PEEP: de 3 em PEEP 4 cmH2O a 10 em PEEP 20 cmH2O. Cinco linhas H foram o limiar para a detecção de hiperdistensão pulmonar, definida como hiperdistensão na TIE >= 24,5%. A área sob a curva ROC foi 0,947 (IC 95% 0.901-0.976). Conclusão: O LUS intraoperatório detectou hiperdistensão pulmonar em valores descendentes de PEEP. A presença de cinco ou mais linhas H podem ser consideradas como indicando hiperdistensão pulmonar
Título en inglés
Ideal PEEP and evaluation of pulmonary atelectasia with the use of ultrasonography during intraoperatory of elective surgeries
Palabras clave en inglés
H lines
Lung/imaging diagnosis
Perioperative care
Positive-pressure respiration
Pulmonary atelectasis
Pulmonary hyperinflation
Pulmonary ventilation
Recruitment maneuver
Resumen en inglés
Purpose: Intraoperative atelectasis occurs immediately after anaesthetic induction and can be detected by lung ultrasound (LUS). However, LUS is considered as unable to assess pulmonary hyperinflation. In this study, we propose a method to detect pulmonary hyperinflation using LUS. Electrical impedance tomography (EIT) was the reference method. Methods: We included 18 patients, 63 ± 6-year old, with normal lungs, undergoing lower abdominal surgery. The following protocol was used: EIT was calibrated, followed by anaesthetic induction, intubation and mechanical ventilation. To reverse posterior atelectasis, a recruitment maneuver - positive end-expiratory pressure (PEEP) 20 cmH20 and plateau airway pressure 40 cmH2O during 120 sec was performed. PEEP titration was then obtained during a descending trial: 20, 18, 16, 14,12,10, 8, 6 and 4 cmH2O. Ultrasound and EIT data were collected at each PEEP level and analyzed by two independent observers. The number of H lines was counted using a special filter. Spearman correlation test and ROC curve were used to compare LUS and EIT data. Results: The number of H lines increased linearly with PEEP: from 3 at PEEP 4 cmH2O to 10 at PEEP 20 cmH2O. Five H lines was the threshold for detecting pulmonary hyperinflation, defined as a mean decrease in maximum EIT compliance >= 24,5 %. The area under the ROC curve was 0.947 (CI 95% 0.901-0.976). Conclusion: Intraoperative transthoracic LUS can detect pulmonary hyperinflation during a PEEP descending trial. Five or more H lines can be considered as indicating pulmonary hyperinflation in normally aerated lung regions
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Fecha de Publicación
2019-03-01
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.