Doctoral Thesis
DOI
https://doi.org/10.11606/T.5.2005.tde-03102005-100300
Document
Author
Full name
Carolina Fu
E-mail
Institute/School/College
Knowledge Area
Date of Defense
Published
São Paulo, 2005
Supervisor
Committee
Carvalho, Carlos Roberto Ribeiro de (President)
Okamoto, Valdelis Novis
Park, Marcelo
Tucci, Mauro Roberto
Vieira, Joaquim Edson
Title in Portuguese
"Comparação entre dois geradores de fluxo e um ventilador específico para ventilação não invasiva no modo CPAP"
Keywords in Portuguese
MODELOS BIOLÓGICOS
PRESSÃO POSITIVA CONTÍNUA NAS VIAS AÉREAS/utilização
RESPIRAÇÃO ARTIFICIAL
Abstract in Portuguese
A pressão positiva contínua nas vias aéreas(CPAP) é a forma mais simples de ventilação não invasiva. O objetivo deste estudo é comparar o desempenho de dois modelos de gerador de fluxo com um ventilador projetado especificamente para ventilação não invasiva no modo CPAP, através de um modelo mecânico de pulmão e em voluntários sadios, em três diferentes alimentações de pressão e em CPAP de 5, 10 e 15 cm H2O. Conclui-se que os geradores de fluxo têm capacidade similar ao ventilador específico para manter o nível de CPAP, se ajustados para sua melhor pressão de alimentação, que foi ao redor de 100 L/min
Title in English
Comparison of two flow generators and a noninvasive ventilator in deliver CPAP
Keywords in English
CONTINUOUS POSITIVE AIRWAY PRESSURE/utilization
MODELS BIOLOGICAL
RESPIRATION ARTIFICIAL
Abstract in English
Continuous positive airway pressure (CPAP) is the simplest form of noninvasive ventilation. Mechanical ventilators designed to invasive ventilation, ventilators specifically designed to noninvasive ventilation (NIV-Ventilator) and continuous flow generators can provide CPAP. The objectives of this study is compare the performance of two continuous flow generators with a ventilator designed to NIV to deliver continuous positive airway pressure (CPAP) and compare flow generators using different oxygen pressure supplies in lung model and healthy subjects, in CPAP of 5, 10 and 15 cm H2O. It was observed that flow generators had similar performance to noninvasive ventilator if they were adjust in their best output flow, that was around 100 L/min
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
CarolinaFu.pdf (616.52 Kbytes)
Publishing Date
2005-10-05
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.