• JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
 
  Bookmark and Share
 
 
Doctoral Thesis
DOI
https://doi.org/10.11606/T.7.2014.tde-04112014-143825
Document
Author
Full name
Rafael Queiroz de Souza
E-mail
Institute/School/College
Knowledge Area
Date of Defense
Published
São Paulo, 2014
Supervisor
Committee
Graziano, Kazuko Uchikawa (President)
Cruz, Aurea Silveira
Moriya, Giovana Abrahão de Araujo
Pinto, Terezinha de Jesus Andreoli
Silva, Luiz Carlos da Fonseca e
Title in Portuguese
Avaliação in vitro da citotoxicidade de instrumentos cirúrgicos enxaguados com diferentes qualidades de água.
Keywords in Portuguese
Água
Centro de Material e Esterilização
Enfermagem
Materiais biocompatíveis
Oftalmologia
Abstract in Portuguese
Introdução: O Centro de Material e Esterilização (CME) deve garantir a remoção completa de resíduos orgânicos e inorgânicos, sobretudo detergentes e endotoxinas, dos produtos críticos para saúde por meio do enxágue suficiente com água de qualidade. Com base nos recentes relatos de síndromes tóxicas do segmento anterior ocular, que foram associadas às más práticas de processamento do instrumental de oftalmologia, especialmente, a qualidade da água utilizada em seu enxágue final, esta Tese teceu a hipótese de que quanto mais pura for a qualidade da água do último enxágue, menor será a citotoxicidade encontrada nos produtos para saúde. Objetivo: Avaliar a citotoxicidade de cânulas de hidrodissecção submetidas à contaminação desafio, limpeza automatizada com enxágue final em diferentes qualidades de água: de torneira, tratada por deionizador, por destilação, por osmose reversa e por ultrapurificação. Métodos: 130 cânulas de hidrodissecção foram usadas, 26 por grupo experimental, caracterizados, de acordo com a água utilizada no último enxágue. As amostras foram submetidas à contaminação desafio interna e externamente por uma solução contendo 20% sangue de carneiro desfibrinado e 80% de Cloreto de Sódio a 0,9%, para simular um cenário desafiador de sujidade no material cirúrgico oftalmológico, sendo em seguida processadas, de acordo com um procedimento operacional padrão validado para este tipo de cânula, variando-se apenas a qualidade de água no enxágue final. Como controle positivo, foram utilizadas três cânulas descartáveis, que foram submetidas ao mesmo procedimento de contaminação desafio, imersas em uma solução contendo água de torneira e detergente enzimático que, após a evaporação da solução, foram esterilizadas. Para o controle negativo, foram usadas três cânulas das mesmas utilizadas nos grupos experimentais, com o mínimo de manipulação e na forma como foram fornecidas pelo fabricante, tendo sido apenas esterilizadas. Adicionalmente, foi incluído um grupo comparativo, no qual três amostras foram processadas em condições idênticas às dos grupos experimentais, enxaguadas com água de torneira, porém, sem secagem. Sequencialmente, procedeu-se a realização do teste de citotoxicidade pela incorporação do corante vital vermelho neutro para a obtenção da média da viabilidade celular de cada grupo experimental. Resultados: Ausência de citotoxicidade nos extratos das amostras, independente da diluição considerada na análise, sendo 83% a porcentagem média de viabilidade celular mínima obtida no extrato das amostras enxaguadas com água tratada por destilação na diluição de 12,5%. Conclusão: As cânulas de hidrodissecção não demonstraram citotoxicidade, independentemente da qualidade de água utilizada no último enxágue. Os resultados apresentados puderam ser alcançados unicamente por meio do uso de um procedimento operacional padrão de limpeza validado, baseado em literatura científica, em recomendações oficiais e na legislação relacionada.
Title in English
Assessment of in vitro cytotoxicity of surgical instruments rinsed with different qualities of water
Keywords in English
Biocompatible Materials
Central Sterile Supply Department
Nursing
Ophthalmology
Water
Abstract in English
Introduction: The Central Sterile Supply Department (CSSD) must guarantee the complete removal of organic and inorganic waste, especially detergents and endotoxins, from critical medical devices through sufficient rinsing with good-quality water. Based on recent reports of toxic anterior segment syndromes, which were associated with poor processing practices of ophthalmology instruments, particularly the quality of water used in the final rinse, this thesis has raised the hypothesis that the better the quality of the water used in the last rinse, the lower the cytotoxicity found in medical devices. Objective: To evaluate the cytotoxicity of hydrodissection cannulas, that were submitted to contamination challenge and then automated cleaning with final rinse in different qualities of water: tap water, water treated by a deionizer device, by distillation, by reverse osmosis and by ultrapurification. Methods: 130 hydrodissection cannulas were used, 26 per experimental group, characterized according to the last rinse water used. The samples were submitted to internal and external contamination challenge using a solution containing 20% defibrinated sheep blood and 80% Sodium Chloride at 0.9%, to simulate a challenging scenario of soiled ophthalmic surgical material, which was then processed, according to a validated standard operating procedure for this type of cannula, changing only the quality of water at the final rinse. Three disposable cannulas were used as positive controls, being submitted to the same contamination challenge procedure, then immersed in a solution containing tap water and enzymatic detergent, and then sterilized after evaporation of the solution. Three cannulas, identical to those used in the experimental groups, were used as negative controls, after minimum handling and exactly as they were supplied by the manufacturer, having only been sterilized. Additionally, a comparison group was included, from which three samples were processed under conditions identical to the experimental groups, rinsed with tap water, but without drying. Sequentially, the cytotoxicity assay was performed by incorporating the neutral red vital dye to obtain mean cell viability from each experimental group. Results: We observed absence of cytotoxicity in the sample extracts, regardless of the dilution considered in the analysis, with 83% of minimal cell viability obtained in the sample extract rinsed with distilled water at 12.5% dilution. Conclusion: Hydrodissection cannulas showed no cytotoxicity, regardless of the quality of water used in the final rinse. The results shown here were achieved by using a validated cleaning standard operating procedure based on scientific literature, official recommendations and related legislation.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
Publishing Date
2014-11-10
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.