  Bookmark and Share
 
 
Disertación de Maestría
DOI
https://doi.org/10.11606/D.75.2018.tde-23102018-170956
Documento
Autor
Nombre completo
João Victor Basolli Borsatto
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
São Carlos, 2018
Director
Tribunal
Lanças, Fernando Mauro (Presidente)
Coltro, Wendell Karlos Tomazelli
Gomes, Paulo Clairmont Feitosa de Lima
Nassur, Maria Eugenia Queiroz
Título en portugués
Colunas empacotadas em cromatografia líquida capilar: desenvolvimento de hardwares e avaliação de suas contribuições no desempenho cromatográfico
Palabras clave en portugués
alargamento de banda
colunas capilares
colunas empacotadas
hardwares para colunas
partículas superficialmente porosas
partículas totalmente porosas
Resumen en portugués

Essa dissertação de mestrado descreve o desenvolvimento de hardwares de colunas para cromatografia líquida capilar. O processo de desenvolvimentos desses dispositivos é descrito gradativamente e os pontos fortes e as limitações de cada modelo de hardware são discutidos. O melhor modelo de hardware desenvolvido apresentou produção simples, fácil conexão ao sistema cromatográfico e resistência a pressões superiores a 900 bar. Sucessivamente ao estabelecimento de um modelo de hardware apropriado, os efeitos dos materiais do hardware na eficiência das colunas foram avaliados. Poucos estudos relatam a influência do hardware nas separações em escala capilar, de forma que essa dissertação contribui para o preenchimento dessa lacuna. Capilares de aço inoxidável e sílica fundida e frits de aço inoxidável e fibra de vidro foram avaliados. Colunas com eficiências superiores a 100.000 pratos por metro foram produzidas.

Título en inglés
Packed columns for capillary liquid chromatography: hardware development and evaluation of its contribution to the chromatography performance
Palabras clave en inglés
band broadening
capillary columns
columns hardware
fully porous particles
packed columns
superficially porous particles
Resumen en inglés
This master's dissertation describes the development of hardware for capillary liquid chromatography columns. The development process of the devices is described gradually and the strengths and limitations of each model of hardware are discussed. The best-developed hardware model presented easy production, practice connection to the chromatographic system and resistance to pressures greater than 900 bar. After the establishment of an appropriate hardware model, the effects of hardware materials on the efficiency of the columns were evaluated. Few studies report the influence of the hardware on capillary scale separations; therefore, this dissertation contributes to fill this gap. Capillaries of stainless steel and fused silica and frits of stainless steel and glass fiber were evaluated. Columns with efficiencies greater than 100,000 plates per meter were produced.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Fecha de Publicación
2018-10-25
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.