Thèse de Doctorat
DOI
https://doi.org/10.11606/T.76.2005.tde-12032005-182218
Document
Auteur
Nom complet
Carlos Ernesto Garrido Salmón
Adresse Mail
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
São Carlos, 2005
Directeur
Jury
Tannus, Alberto (Président)
Araújo, Dráulio Barros de
Engelsberg, Mario
Magon, Claudio Jose
Rabbani, Said Rahnamaye
Titre en portugais
"Desenvolvimento e aplicações clínicas de um sistema integrado transdutor/bobinas de gradientes de alto desempenho para obtenção de imagens por ressonância magnética em 0.5 TESLA"
Mots-clés en portugais
bobinas de gradientes
Engenharia biomédica
Imagens por Ressonância Magnética
transdutor de RF
Resumé en portugais
Este trabalho descreve o desenvolvimento de um sistema integrado transdutor/bobinas de gradientes de alto desempenho para Imagens por Ressonância Magnética. Este sistema é composto por um transdutor de radiofreqüência tipo sela e um conjunto de 3 bobinas locais assimétricas. No desenho do transdutor foram otimizados os parâmetros: relação sinal ruído e uniformidade do campo magnético por ele gerado. A densidade de corrente de cada bobinas local foi otimizada mediante técnicas numéricas estocásticas para gerar um gradiente de campo magnético uniforme em cada uma das 3 direções do espaço numa região das dimensões do crânio. O conjunto de bobinas de gradientes construído possui um diâmetro livre de 31.5 cm e gera em média 25 mT/m/A por bobina, com indutâncias inferiores a 310 mH. São mostradas as aplicações clínicas desenvolvidas nas áreas de imagens tridimensionais e angiografia, a partir das seqüências de pulsos implementadas e aproveitando o uso do sistema integrado, em um tomógrafo de ressonância magnética de 0.5 Tesla. Imagens de phantom foram adquiridas em menos de 500 ms usando o conjunto integrado e técnicas do tipo Echo Planar Imaging. Aspectos referentes à caracterização e correção de campos magnéticos estáticos e homogêneos são também comentados. As soluções descritas nesta tese têm um amplo conteúdo tecnológico e beiram nas fronteiras da Física Aplicada e a Engenharia Biomédica.
Titre en anglais
"Development and clinical applications of a high performance radio-frequency/gradient coil integrated system for Magnetic Resonance Imaging in 0.5 Tesla"
Mots-clés en anglais
Angiography
Biomedical Engineering
gradient coils
MRI
RF coil
SSFP
Resumé en anglais
Here we describe the development of a high performance radio-frequency/gradient coil integrated system for Magnetic Resonance Imaging. A saddle radio-frequency coil and a three-axis asymmetric local gradient coil composed this system. Two parameters were optimized in the RF coil design: signal-to-noise ratio and magnetic field uniformity. The current density of each local coil was optimized using stochastic numerical techniques, in order to generate a uniform magnetic field gradient by axis in a region representing a human head. The build gradient coil set has an inner diameter of 31 cm. The average gradient efficient of the three-axis is 25 mT/m/A and the maximum inductance is less than 310 mH. We show the clinical applications performed in three-dimensional and angiography imaging areas in a 0.5 Tesla magnetic resonance tomograph. These applications were optimized to taking advantage from the integrated system. Phantom images were acquired in less than 500 millisecond using echo planar techniques and the integrated set. Some aspects about static and homogeneous magnetic field characterization and correction are also commented. In this work we described solutions with wide technologic content close to the boundaries of the Applied Physics and Biomedical Engineering.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
tesecarlosgarrido.pdf (4.95 Mbytes)
Date de Publication
2005-04-29
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.