Mémoire de Maîtrise
DOI
https://doi.org/10.11606/D.76.2004.tde-28042004-184322
Document
Auteur
Nom complet
Fernando Fernandes Paiva
Adresse Mail
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
São Carlos, 2004
Directeur
Jury
Tannus, Alberto (Président)
Rabbani, Said Rahnamaye
Schneider, Jose Fabian
Titre en portugais
"Metodologia de imagens de NMR que utiliza um único pulso adiabático de inversão."
Mots-clés en portugais
Pulsos Adiabáticos
Ressonância Magnética Nuclear
Resumé en portugais
Baseado em uma seqüência já conhecida de Multislice Spin Echo convencional, implementamos uma seqüência do tipo Modified Driven Equilibrium Fourier Transform (MDEFT) adiabática multislice em um sistema de 2T e um de microimagens (9.4T). Essa metodologia utiliza, como pulso de inversão, um pulso de RF com modulação em amplitude e freqüência, conhecido por pulso adiabático. Dessa forma, os tempos de inversão são iguais para todos os planos selecionados, resultando num contraste por T1 uniformemente distribuído ao longo dos mesmos. O uso de um único e longo pulso adiabático de inversão tem, ainda, a vantagem de utilizar uma potência de pico menor para operar da mesma forma que um grupo de m pulsos convencionais. Utilizamos pulsos cujas modulações correspondentes de fase e freqüência foram calculadas com base nas condições de “Offset Independent Adiabaticity" (OIA) e a real eficiência dos mesmos foi avaliada através de simulações e experimentalmente. Para realização dos testes da metodologia desenvolvemos e caracterizamos dois “phantoms", cujas imagens demonstram a aplicabilidade da metodologia nos dois sistemas utilizados.
Titre en anglais
"NMR imaging methodology using a single adiabatic inversion pulse."
Mots-clés en anglais
Adiabatic Pulses
Nuclear Magnetic Resonance
Resumé en anglais
Based on the conventional Multislice Spin Echo pulse sequence, we implemented an adiabatic multislice Modified Driven Equilibrium Fourier Transform (MDEFT) pulse sequence on a 2T and a 9.4T microimaging system. This methodology uses, as inversion pulse, an amplitude and frequency modulated RF pulse, known as adiabatic pulse. In that way, the inversion times are the same for all slices, resulting in an evenly distributed T1 contrast. A single and long adiabatic inversion pulse has in addition the advantage of using a reduced peak power to perform in the same way as the group of m conventional pulses. We used pulses whose corresponding phase and frequency modulations were based on Offset Independent Adiabaticity and their real efficiency were evaluated both through simulations and experimentally. To test the methodology, we developed and characterized two phantoms, whose images demonstrate the applicability of the methodology in the two mentioned systems.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
Dissertfernando.pdf (2.93 Mbytes)
Date de Publication
2004-08-16
 