Master's Dissertation
DOI
https://doi.org/10.11606/D.85.2010.tde-01062011-122808
Document
Author
Full name
Daniel Menezes Dias
E-mail
Institute/School/College
Knowledge Area
Date of Defense
Published
São Paulo, 2010
Supervisor
Committee
Potiens, Maria da Penha Albuquerque (President)
Nersissian, Denise Yanikian
Vivolo, Vitor
Title in Portuguese
Estabelecimento de um novo método de calibração de câmaras de ionização tipo lápis para dosimetria em feixes de tomografia computadorizada
Keywords in Portuguese
câmara de ionização tipo lápis
dosimetria em tomografia computadorizada
metrologia em CT
radiodiagnóstico
Abstract in Portuguese
As câmaras de ionização tipo lápis são utilizadas para dosimetria de feixes em equipamentos de tomografia computadorizada (CT). Neste estudo foi estabelecida uma nova metodologia de calibração destas câmaras, visando a adequação do Laboratório de Calibração de Instrumentos do Instituto de Pesquisas Energéticas e Nucleares (LCI) aos atuais padrões metrológicos internacionais que tratam dos procedimentos específicos para calibração das câmaras utilizadas em CT. Primeiramente partiu-se da implantação das novas qualidades de radiação padrão RQT, que simulam feixes não atenuados usados para aplicações em CT, de acordo com a IEC 61267 da International Electrotechnical Commission (IEC). Após o estabelecimento dos campos padrões de radiação, foi implantada a metodologia de calibração específica para câmaras lápis segundo o Technical Report Series No. 457, da International Atomic Energy Agency (IAEA), que descreve as particularidades do procedimento a ser seguido pelos Laboratórios de Dosimetria Padrão Secundário (SSDLs) em termos de colimação e posicionamento em relação ao feixe de radiação. Inicialmente, foram realizadas medições de PPV (kV) e determinação das filtrações adicionais de cobre por meio dos valores de camada semi-redutora (CSR) recomendados pela norma IEC 61267, para que as qualidades de radiação de referência: RQT 8, RQT 9 e RQT 10 fossem implantadas. Para os filtros adicionais, foram utilizados alumínio e cobre de alta pureza (maior ou igual a 99,9%). Uma metodologia alternativa para determinar as filtrações adicionais também foi utilizada, através da qual se encontrou diretamente as RQTs em espessuras de filtros de cobre equivalente ao conjunto RQR (Al) + Filtração Adicional (Cu), sendo esta uma opção quando não se têm implantadas as RQRs. Com a implantação desta nova metodologia de calibração das câmaras de ionização tipo lápis, o LCI está pronto para calibrar estes instrumentos de acordo com as recentes recomendações internacionais. Permitindo assim uma melhoria na rastreabilidade das calibrações, como também na qualidade dos serviços metrológicos que o IPEN oferece para todo o País.
Title in English
Establishment of a new calibration method of pencil ionization chamber for dosimetry in computed tomography
Keywords in English
dosimetry in computed tomography
pencil ionization chamber
Abstract in English
Pencil ionization chambers are used for beam dosimetry in computed tomography equipment (CT). In this study, a new calibration methodology was established, in order to make the Calibration Laboratory of Instituto de Pesquisas Energéticas e Nucleares (LCI) suitable to international metrological standards, dealing with specific procedures for calibration of these chambers used in CT. Firstly, the setup for the new RQT radiation qualities was mounted, in agreement with IEC61267 from the International Electrotechnical Commission (IEC). After the establishment of these radiation qualities, a specific calibration methodology for pencil ionization chambers was set, according to Technical Report Series No. 457, from the International Atomic Energy Agency (IAEA), which describes particularities of the procedure to be followed by the Secondary Standard Dosimetry Laboratories (SSDLs), concerning to collimation and positioning related to the radiation beam. Initially, PPV (kV) measurements and the determination of copper additional filtrations were carried out, measuring the half value layers (HVL) recommended by the IEC 61267 standard, after that the RQT 8, RQT 9 and RQT 10 radiation quality references were established. For additional filters, aluminum and copper of high purity (around 99.9%) were used. RQTs in thickness of copper filters equivalent to the set RQR (Al) + Additional Filtration (Cu) was directly found by an alternative methodology used to determine additional filtrations, which is a good option when RQRs have not the possibility of be setting up. With the establishment of this new methodology for the ionization pencil chambers calibration, the LCI is ready to calibrate these instruments according to the most recent international standards. Therefore, an improvement in calibration traceability, as well as in metrological services offered by IPEN to all Brazil is achieved.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
Publishing Date
2011-06-21
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.