Doctoral Thesis
DOI
https://doi.org/10.11606/T.85.2017.tde-20092017-155553
Document
Author
Full name
Felipe Belonsi de Cintra
E-mail
Institute/School/College
Knowledge Area
Date of Defense
Published
São Paulo, 2016
Supervisor
Committee
Caldas, Linda Viola Ehlin (President)
Medeiros, Regina Bitelli
Moralles, Mauricio
Pinto, Teresa Cristina Nathan Outeiro
Vivolo, Vitor
Title in Portuguese
Desenvolvimento de um calorímetro com núcleo de água e de uma unidade de tratamento de líquidos para dosimetria de radiação gama
Keywords in Portuguese
calorímetro de água
calorímetros
dosimetria de radiação gama
Abstract in Portuguese
Este trabalho consiste no desenvolvimento de um detector composto por um calorímetro com núcleo de água para dosimetria de feixes de 60Co e 137Cs com média intensidade. Além do calorímetro, foi também dimensionada a instrumentação, a metodologia de calibração e a medição de temperatura. O trabalho foi dividido em 4 etapas distintas: escolha do melhor projeto como base para desenvolvimento do calorímetro, dimensionamento de parâmetros e ajuste fino do projeto, construção e calibração. Durante o desenvolvimento deste trabalho foi construída uma Unidade de Tratamento de Água que permite a produção da água que o calorímetro emprega. Em seguida foi construído o calorímetro em sí com os conceitos das etapas anteriores. O equipamento foi testado tanto nas instalações da GMR quanto no CTR do IPEN, obtendo taxas de dose na água com fontes de 137Cs e 60Co com diversas atividades (3,3 TBq, 15,54 TBq e 7 TBq) e em algumas distâncias. Foram obtidas taxas de dose que variavam de 2 mGy/s até 15 mGy/s, dependendo do tipo de fonte e SDD escolhida. O projeto contou com simulações de códigos como MCNP5 e FLUENT 14 e foram essenciais na construção do equipamento. O objetivo deste trabalho foi aprimorar a infraestrutura do LCI, desenvolvendo um detector que deverá aumentar a gama dos serviços prestados pelo LCI com um tipo de medição dosimétrica ainda inexistente no país.
Title in English
Development of a water calorimeter and a liquid treatment unit for gamma radiation dosimetry
Keywords in English
calorimeters
gamma dosimetry
water calorimeter
Abstract in English
This work aims to develop a radiation detector composed by a water calorimeter for beam dosimetry of 60Co and 137Cs gamma radiation with medium intensity. Besides the calorimeter itself, all the instrumentation, methodology, calibration and testing protocols were established. This work was divided into 4 parts: the choice of the best design; simulation and setting of the ideal operation parameters; adaptation of the project; construction and calibration. During the development of this work a water treatment unit was built which allows the production of distilled water for the calorimeter. Then the calorimeter itself was built with the concepts from the previous steps. The equipment has been tested both in the facilities of GMR as the CTR at IPEN, obtaining dose rates ( from 2 mGy/s up to 15 mGy/s) in water for 137Cs and 60Co sources in some activities (3,3 TBq, 15,54 TBq e 7 TBq) and distances. The MCNP5 and FLUENT 14 have been used and proved essential in the development of the prototype. The main goal was to improve the LCI infrastructure with a new radiation detector that will increase the calibration laboratory service range with a still absent kind of dosimetric measurement in Brazil.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.