Thèse de Doctorat
DOI
https://doi.org/10.11606/T.85.2014.tde-29092015-083645
Document
Auteur
Nom complet
Luís Cláudio Aranha
Adresse Mail
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
São Paulo, 2014
Directeur
Jury
Rossi, Jesualdo Luiz (Président)
Bonachela, Wellington Cardoso
Braga, Francisco Jose Correa
Casemiro, Luciana Assirati
Mucsi, Cristiano Stefano
Titre en portugais
Estudo dos parâmetros de estabilidade de travamento entre implante dentário e componente protético através do sistema cone Morse
Mots-clés en portugais
cone Morse
contaminação
fadiga
implante
microtorque
Resumé en portugais
Na implantodontia visa-se dois grandes resultados para os desdentados parciais e totais, função mastigatória e estética. Ambos têm o mesmo grau de importância, para tanto se deve pensar no que se tem de mais seguro e prático nos dias de hoje. O cone Morse, na implantodontia oral, é um sistema de estabilidade entre o implante dentário e o parafuso intraósseo de sustentação. Para seu correto uso se faz necessário uma série de análise de suas características e condições. No presente trabalho focou-se em desenvolver protocolo de validação de torque, além de analisar parâmetros de acabamento, deformação e estabilidade précarga e pós-carga dentro de protocolos já recomendados. Avaliou-se sua condição antes e após os efeitos de carga mastigatória simulada por ensaio de fadiga. Concluiu-se que o uso do cone Morse dentro dos protocolos indicados deve ser revisto quanto a padrões similares de torque nos pilares sólidos e indexados, cada um possui limites e indicações diferentes para se manter estável. Concluiu-se também que não é um sistema totalmente livre de trânsito de bactérias e que esse trânsito aumenta de forma proporcional ao uso e força mastigatória, indicando que melhorias no acabamento entre as peças devem ser implementadas e/ou adição de material que afaste crescimento bacteriano priorizando a saúde do periodonto e do tecido ósseo de sustentação mantendo o sistema seguro em longevidade.
Titre en anglais
Study on the stability parameters of locking between dental implants and prosthetic component system through the Morse taper
Mots-clés en anglais
contamination
fatigue
implant
microtorque
Morse taper
Resumé en anglais
In implantology it is aimed two major outcomes for the partial and total toothless persons, mastigatory function and aesthetics. Both have the same degree of importance, therefore it must be thought about what is safer and more practical these days. Morse taper in oral implantology is a system of stability between the dental implant and the intraosseous screw support. For its correct use a series of analysis of its features and conditions is required. This work was focused on developing torque validation protocol, and analyzing surface finishing parameters, deformation and stability as before load and after load in recommended protocols. The implants condition before and after the effects of habitual masticatory load was simulated by fatigue testing. It was concluded that the Morse taper use within the specified protocols should be reviewed for similar patterns of torque in solids and indexed pillars, each has different limits and signs to remain stable. It was also concluded that it is not a traffic totally free system of bacteria and that this influence increases in proportion to the use and masticatory forces, improvements in the surface finish between parts and / or addition of material that hinders bacterial growth is needed if it is prioritized the health of the periodontium and the bone tissue keeping the system safe in the long therm.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
2014AranhaEstudo.pdf (3.42 Mbytes)
Date de Publication
2015-10-09
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.