Título en portugués
Calibração indireta de máquinas de medir a três coordenadas
Palabras clave en portugués
calibração
máquinas de medir a três coordenadas
transformação homogênea
Resumen en portugués
As máquinas de medir a três coordenadas (MM3Cs) possuem erros inerentes a sua estrutura que afetam a acuracidade e a repetibilidade das medições. Dos erros presentes nestas máquinas, os erros geométricos são, na maioria das vezes, os de maior influência. Para caracterizar esses erros procedimentos de avaliação das MM3Cs fazem uso de processos de calibração direta e indireta. A calibração indireta utiliza artefatos padrões pré-qualificados e normalizados, considerados mais rápidos e econômicos. Este trabalho tem por objetivo calibrar indiretamente uma MM3C do tipo "Ponte Móvel" utilizando uma barra de esferas e seus resultados comparados com o método da sintetização de erros. Envolve a medição de uma barra de esferas em 20 posições e orientações diferentes no volume de trabalho da máquina, conforme a norma ANSI/ASME B 89 4.1, 1995. A distância entre os centros das esferas foi calculada e comparada à distância padrão determinando o erro da máquina nas posições em que a Barra de Esferas foi medida. Posteriormente, as coordenadas dos centros das esferas foram sintetizadas no volume de trabalho da MM3C utilizando as equações de sintetização que foram obtidas através do uso de técnicas de tranformação homogêneas. Os resultados obtidos foram comparados, demonstrando que ambos os métodos apresentam resultados similares, porém, a calibração utilizando uma barra de esferas é mais rápida e econômica. Por sua parte o método de sintetização de erros necessita de equipamentos mais sofisticados
Título en inglés
Indirect calibration of three coordinates measuring machines
Palabras clave en inglés
calibration
homogenous transformation
three coordinate measuríng machines
Resumen en inglés
The measurement accuracy and the repeatability of Three Coordinates Measuring Machines (3CMM) are affected by several errors sources. Those CMM errors are characterized trough calibration procedures fhat can be divided in direct and indirect calibration methods. The indirect calibration uses standards artifacts as reference and usually it is faster and more economic than the direct calibration. The two direct methods are known as synthetization method and the grid method. They have as main objective the determination of the machine volumetric en-ors. In this work a "Moving Bridge" CMM was indirectly calibrated using as reference Bali Bars. The results of this calibration were compareci wifh fhose obtained with the synthetization method. The Bali Bars was measured in 20 (twenty) positions and orientations within the machine working volume. The measurement positions are recommended by the ANSI / ASME B89 4. l. The distance between the centers of the spheres of Bali Bars was calculated and compareci with the calibrated distance of the Bali Bars. Ïn this way, the machine error for each measured position was determinecL F\.irther, fhe volumetric error of center coordinates of spheres were determíned using the errors sinthetization equaüons obtained wifh. the application of homogenous transformation techniques. The comparison showed that the results of both calibrations method are veiy similar. The indirect calibration wifh the Bali Bars is, however, cheap and faster than the direct calibration fhat needs sophisticated measurement equipment.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Fecha de Publicación
2024-06-03
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.